DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
8 - 72 Section 8-9 - Software Loading Procedure - Software Version v2.x
8-9-14-2 Vivid 7 System Software Loading - Software Version v2.x (cont’d)
5.) When requested, enter the missing part of the Computer Name (the unit’s serial number) as you
recorded it in Table 8-16 "Record settings from Tcpip screen (v2.x)" on page 8-63.
Example: VIVID7-001890
Always enter two zeros before you enter the serial number. This means that when you have a serial
number which is 1890 you enter 001890. When the serial number is 1123 you enter 001123.
6.) In this step, you can choose if you want to do a complete Vivid 7 installation or if you only want to
update or reinstall the software on the C:\ partition (C:\ drive).
Usually it should be OK to select “B” to update the C-drive only. This will keep Presets etc.
a.) Press B to update/re-install the software on the C:\ partition. (Only data on the C:\ partition will
be erased. System configurations etc. will be kept.) Continue with step 7.
b.) Use ONLY if a “B” update cannot be used: Press A to do a complete Vivid 7 software
installation. (All data will be erased.)
c.) Press C to Exit.
Figure 8-67 Serial Number entry example
NOTICE
If you select “A” in the next step, ALL existing software and data on the hard disk will be erased. If
backup has not been performed as described earlier in this manual, all data like Patient Database,
System Configuration and User Configurations (Customer Presets), will be lost.
Figure 8-68 Info entry page
Enter the Serial
Number here
Press A, B or C to
choose the wanted
option.