GE HEALTHCARE
DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
8 - 76 Section 8-9 - Software Loading Procedure - Software Version v2.x
8-9-15-2 Verify Video and VCR Settings - Software Version v2.x
1.) Select Service (lower, right corner)
.
2.) From Video Settings pull down menu, verify Format as recorded in Table 8-19 "Video Settings and
VCR Type" on page 8-68. Correct the selection if needed.
3.) From the VCR pull down menu, verify VCR type as recorded in Table 8-19 on page 8-68. Correct
the selection if needed.
8-9-15-3 Verify TCP/IP Settings - Software Version v2.x
1.) From the top of the connectivity tab, select the Tcpip tab.
2.) Verify that all settings on the TCPIP tab matches those recorded in Table 8-16 "Record settings from
Tcpip screen (v2.x)" on page 8-63.
3.) Select Save settings.
4.) Select Ok to continue.
Figure 8-72 Video Settings
Figure 8-73 TCP/IP Example
Settings