8-9-15 Verifications - Software Version v2.x
NOTE: Software Option Strings are software version dependant. If a new software version has been
installed, please follow installation instructions for that software. Usually, new software option
strings are needed.
8-9-15-1 Verify Software Option Key Settings - Software Version v2.x
1.) Press CONFIG
(F2) and log on as adm, see 4-2-4 "Log On to the System as ‘ADM’" on page 4-12.
2.) Select Admin (lower part of window),
3.) Select the SYSTEM ADMIN tab.
4.) Verify that the SW Options Keys (alphanumeric strings) in the SW Options Key window are the
same as recorded in 8-9-10 "Record SW Option Keys - Software Version v2.x" on page 8-62.
5.) Select New to open the New Key dialog where you can type any missing Software Option Key(s)
(alphanumeric string(s)).
Figure 8-70 System Admin screen
Figure 8-71 Type Software Option Key
System Admin Tab
New
Admin
Field for Software Option Key (Serial String)

Table of Contents