C Error Messages
306 Keysight Models 6811C, 6812C, and 6813C Programming Guide
Device-dependent errors 100 through 32767 (sets Standard Event Status Register bit #3)
200 Outgrd not responding
201 Front panel not responding
210 Ingrd receiver framing error
211 Ingrd uart overrun status
212 Ingrd received bad token
213 Ingrd receiver buffer overrun
214 Ingrd input buffer overrun
215 Outgrd output buffer overrun
216 RS-232 receiver framing error
217 RS-232 receiver parity error
218 RS-232 receiver overrun error
219 Ingrd inbuf count sync error
220 Front panel uart overrun
221 Front panel uart framing
222 Front panel uart parity
223 Front panel buffer overrun
224 Front panel timeout
401 CAL switch prevents calibration
402 CAL password is incorrect
403 CAL not enabled
404 Computed read back cal constants are incorrect
405 Computed programming cal constants are incorrect
406 Incorrect sequence of calibration commands
481 Model bit detection failed
501 LAN disconnect
Table C-1 Error numbers (continued)
Error # Error string [description/explanation/examples]