EasyManuals Logo

LMI Technologies Gocator 2300 Series User Manual

LMI Technologies Gocator 2300 Series
431 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #98 background imageLoading...
Page #98 background image
Gocator 2300 & 2880 Series
Gocator Web Interface • Scan Setup and Alignment • 98
To configure X or Ygap filling:
1. Go to the Scan page.
2. Choose Profile or Surface mode in the Scan Mode panel.
If one of these modes is not selected, you will not be able to configure gap filling.
3. Expand the Filters panel by clicking on the panel header or the button.
4. Click on the Gap Filling tab.
5. Enable the X or Y setting and select the maximum width value.
The value represents the maximum gap width that the Gocator will fill. Gaps wider than the maximum
width will not be filled.
6. Save the job in the Toolbar by clicking the Save button .
7. Check that the laser profiling is satisfactory.
Median
The Median filter substitutes the value of a data point with the median calculated within a specified
window around the data point.
Missing data points will not be filled with the median value calculated from data points in the
neighbourhood.
To configure X or Ymedian:
1. Go to the Scan page.
2. Choose Profile or Surface mode in the Scan Mode panel.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LMI Technologies Gocator 2300 Series and is the answer not in the manual?

LMI Technologies Gocator 2300 Series Specifications

General IconGeneral
BrandLMI Technologies
ModelGocator 2300 Series
CategoryAccessories
LanguageEnglish

Related product manuals