— Corner
correction
Platform 1, platform 2 (only shown if the
number of load cells is = 8).
O.k., when realized.
— Modify May be used only for minor changes (e.g.
changing the dead load). Otherwise, always use
[New]!
— Param Parameter settings
— Ambient
conditions
Select environmental conditions: Very stable,
stable, unstable, very unstable
— W&M Selection legal-for-trade mode: <none>, OIML,
NSC, NTEP
— Unbal.
Check deviat.
Unbalanced check deviation: The plausibility
check is activated when the average deviation
is >0%.
Input: 0…100 %
— Standstill time Input of the standstill period:
0.01 s…<0.50 s>…2.0 s (The range depends on
the measuring time.)
— Standstill
range
Input of the standstill range:
0.00 d…<1.00 d>…10.00 d (The range depends
on the measuring time.)
— Tare timeout Input of timeout when there is no standstill:
0.1 s…<2.5 s>…25 s
— Zeroset range ± Range of zero point when there is no
standstill. Input:
0.00 d…<50.00 d>…10000.00 d
— Zerotrack
indic. range
Input of the zerotrack indication range:
0.00 d…<0.25 d>…10000.00 d
— Zerotrack step Input of the zerotrack step:
0.00 d…<0.25 d>…10.00 d
— Zerotrack time Input zerotrack time: <0.0 s>…25 s
— Overload Input of the weighing range above the
maximum load (Max) without error message:
0…<9 d>…999999 d
— Min Input of the minimum load:
0 d…<50 d>…999999 d
— Range mode Range selection: <Single range>, Multiple
range, Multi-interval
See also Chapters 7.14.15.2 and 7.14.15.3.
— Range limit 1 Input of the limit 1: In weight, unit same as Max,
transition from small to medium verification
interval
Only for [Multiple range] or [Multi-interval]!
— Range limit 2 Input of the limit 2: In weight, unit same as Max,
transition from small to medium verification
interval
Only for [Multiple range] or [Multi-interval]!
— LC name Give each load cell a name.
— LC 1…n load cell 1…n, e.g.: PR6224-xx
Input of max. 20 alphanumeric characters.
— Default Settings are reset to factory settings.
7 Getting started X3 Process Indicator PR 5410
Minebea Intec EN-159