16 Special Functions
16.6 GC Configuration
254
GC-2014
DAFC UNIT
Selection: DAFC/AMC. LR, Default: DAFC
When using the dual AFC as an AMC for detector makeup gas, set to AMC. LR. If AMC.
LR is set, the injection port connected to the dual AFC becomes off from the flow line
and is not displayed on the flow line configuration screen (8.3.1)
Set the DAFC unit prior to starting GCsolution.
DTCD PREAMPLIFIER
Selection: x1/x10, Default: x1
When it is set to x10, TCD output is amplified ten times.
GC START
Selection: SYSTEM Key Screen/ SET Key Screen/ MONIT Key Screen/,
Default: SYSTEM Key Screen
Select the screen, which is displayed automatically after pressing the "Start GC" key
(PF1 key of the [SYSTEM] key main screen).

Table of Contents

Other manuals for Shimadzu GC-2014

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu GC-2014 and is the answer not in the manual?

Shimadzu GC-2014 Specifications

General IconGeneral
TypeGas Chromatograph
DisplayLCD
Maximum Heating Rate120°C/min
Data AcquisitionDigital
Maximum Number of Detectors4
WeightApprox. 45 kg
Carrier GasHelium, Nitrogen, Hydrogen
Detector OptionsFID, TCD, ECD, FPD
Injection PortSplit/Splitless, On-column
Gas ControlElectronic Flow Control (EFC)
Temperature Programming Rate0.1 to 120 °C/min
Power Requirements230 V AC, 50/60 Hz
Detector Temperature RangeUp to 400°C (depending on detector type)