EasyManua.ls Logo

Applied Biosystems 3500 - Page 124

Applied Biosystems 3500
326 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
spectral calibration using the Z dye set. Do not select Keep Spectral Calibration
Data.
3.
Select the number of w
ells and plate position in the instrument.
Note: You do not create a plate in the software for the install check.
4.
Specify the starting well for the injection position in which you loaded the
standard in the plate.
Note: If you navigate away from the Install Standard screen after you start the
install check, the starting well may be reset to A01. This is a display issue only;
the starting well you specify is used for the install check.
5.
(Optional) If you have not already run a spectral calibration, select Keep Spectral
Calibration Data to save the sequencing install standard run (if it passes) as a
spectral calibration.
Note: The spectral calibration record will only be saved if the Keep Spectral
Calibration Data option is checked on the screen. If you uncheck the option,
create a separate spectral calibration from the Maintenance menu.
Chapter 6 Run c
alibrations and install checks
Run a Sequencing install check
6
124
3500/3500xL Genetic Analyzer User Guide—Data Collection Software v3.1

Table of Contents

Related product manuals