Archive, purge, and restore audit records
The audit archive function makes a copy of audit records. Purge makes a copy of
audit records, and then deletes them. You can use the Restore function to restore
purged audit records.
Archive and purge
To selectively archive or purge (delete) system conguration or action audit records:
1.
Select records in the appropriate screen.
2.
Click
Archive Audit Records or Purge Audit Records.
3.
If y
ou select Archive:
Specify a location and name for the .asz audit archive le.
(Optional) Click Yes to Purge (delete) the records after archive.
Restore
To restore system conguration or action audit records, click
Restore, then select
the .asz le to restore.
Export audit records
As needed, you can export audit records to a .txt le for additional manipulation and
reporting outside the software.
1.
Display the records of interest.
2.
Select the records to export.
3.
Click
Export Audit Records.
4.
Specify a name and location for the export .txt le.
5.
Click Save.
Not
e: If you export audit records for samples that are not in their original
location (samples have been deleted or moved), an error message is displayed.
Return sample data les to their original location, then export again.
Chapter 8 Use Security
, Audit, and E-Sig functions (SAE Module)
Administrators
8
210
3500/3500xL Genetic Analyzer User Guide—Data Collection Software v3.1

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 3500 and is the answer not in the manual?

Applied Biosystems 3500 Specifications

General IconGeneral
Number of capillaries8
Capillary Length50 cm
LaserSolid-state laser
Instrument TypeCapillary electrophoresis system
TechnologyCapillary Electrophoresis (CE)
Weight95 kg
Power Requirements100–240 V

Related product manuals