EasyManuals Logo

Applied Biosystems 3500 User Manual

Applied Biosystems 3500
326 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #265 background imageLoading...
Page #265 background image
Symptom
Possible cause Action
Low signal. Not enough sample: Pipetting error. Prepare new sample.
Sample has high salt concentration. Dilute or desalt samples.
Insufficient mixing. Vortex the sample thoroughly, and
then centrifuge the tube to condense
the sample to the bottom of the tube.
Weak amplification of DNA. Reamplify the DNA.
Check DNA quality.
Sample volume is <10 µL. Check that sample volume is at least
10 µL.
Autosampler out of calibration. Contact Thermo Fisher Scientific.
Elevated baseline. Possible contaminant in the polymer
path.
Run the Wash Pump and Channels
wizard.
Poor spectral calibration. Perform new spectral calibration.
Loss of resolution. Too much sample injected. Dilute the sample and re-inject.
Poor quality water. Use distilled or deionized water.
Degraded polymer. Replace polymer.
Capillary array used for more than
160 injections.
Replace the capillary array. Run the
Install Capillary Array wizard.
Degraded formamide. Prepare fresh HiDi
Formamide (see
“HiDi
Formamide“ on page 22 for
storage conditions) for sample
preparation.
Sample has high salt concentration. Dilute or desalt samples.
Poor resolution in some capillaries. Insufficient filling of capillary array. Tighten the connectors and array
locking lever. Run the Fill Array with
Polymer wizard and look for polymer
leakage. Check for broken capillaries,
run the Install Capillary Array wizard
if needed.
Re-inject the same samples.
Poor quality samples. Check the sample preparation.
Leak in system. Tighten the connectors and array
l
ocking lever.
No current. Not enough buffer in ABC. Ensure that the buffer is filled up to
the fill line. See “Check buffer fill
levels“ on page 34.
Appendix A Troubleshoot
Data/electropherogram troubleshooting
A
3500/3500xL Genetic Analyzer User Guide—Data Collection Software v3.1
265

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 3500 and is the answer not in the manual?

Applied Biosystems 3500 Specifications

General IconGeneral
Number of capillaries8
Capillary Length50 cm
LaserSolid-state laser
Instrument TypeCapillary electrophoresis system
TechnologyCapillary Electrophoresis (CE)
Weight95 kg
Power Requirements100–240 V

Related product manuals