Performing a Dynamic Assessment Test
Note
In the event that during the last 3 minutes the monitoring results were stable, the following
message is displayed providing the user with the option to commence the Bolus Test from the
Challenge stage:
Tap
Yes
to confirm - the Challenge commences immediately (with the last 3 minutes becoming the
Baseline) - see page 5-49.
Otherwise, tap
No
and continue to Step 3 below.
3. As requested in Figure 5-42, instruct the patient to keep still for the next 3 mins, then
tap Next.
During this time, the system will obtain a Baseline for measurement.
While the Baseline measurement is in progress, the Dynamic display shows a graph
being drawn slowly from left to right, as well as a 3-minute countdown (min/secs). See
the example in Figure 5-43.