6-2 Starling Monitor - User Guide
Cleaning Procedures
Cleaning Procedures
Cleaning and Disinfecting the Starling Monitor
The Starling Monitor requires routine cleaning, which includes removal of any soil or dirt from
the external surfaces. A soft cloth dampened lightly with water may be used.
Whenever necessary, disinfect external Monitor surfaces, and reusable accessories (such as
the Starling Patient Cable), by cleaning with 70% medical-grade alcohol using a dampened
soft towel or wipe.
Refer to the precautions and procedure instructions below.
WARNING
Do not perform cleaning procedures when power to the Starling Monitor is ON.
This may be hazardous to the user and/or may cause damage to the system!
CAUTION
To prevent patient-to-patient disease transmission, adequate cleaning and disinfection of the
Starling Monitor and reusable accessories are necessary.
It is the responsibility of the equipment user to verify and maintain the effectiveness of the
infection control procedures in use.
Important
Before performing cleaning or disinfection procedures, ensure power to the Monitor is OFF (see System
Shutdown Procedure on page 4-13), and that the AC power cable is disconnected from the mains power.
CAUTION
Do not expose the Starling Monitor, Starling Patient Cable or Starling Sensors to sprays, liquids or
any other type of solvent.
When performing cleaning procedures, do not use abrasive scouring powders or pads, caustic
detergents, oxide, hypochlorite, lacquer, thinners, ammonia, or acid-based cleaning solutions.
Use of these materials may cause irreparable damage to the Monitor and/or accessories!