CAPINTEC, INC. CAPTUS
®
3000
Figure 16-6
14. In the Measurement Name text box, input a name for the measurement (e.g. Sample
1). (32 characters max)
15. Click the appropriate radio button in the Measurement Type frame to select a
Measurement Type. The default is Sample.
16. Input the Default Sample Volume or Quantity. This field is optional. The value cannot
be entered if the Measurement Type is Background.
17. The Options frame is only active when the Measurement Type is Sample. Check Ask
for Actual Sample Quantity if the user is to input the Sample Quantity for the
measurement. Check Ask for Sample Date / Time if the user is to input the date and
time the sample was collected.
18. In the Counting Time text box, input the counting time in seconds. (10-9,999
seconds)
19. If special instructions are to be displayed for the measurement, type these
instructions in the Instructions for Sample text box.
20. When all the fields are entered, click the
button or press ALT+N to enter
data for the next measurement. For the last measurement, click the
button
or press ALT+D to end the protocol entry.
Note: The number of measurement setup screens that appear depends upon the
number of measurements entered in step 3 above.
16-6 CUSTOM PROTOCOLS November 13