EasyManuals Logo

Capintec CAPTUS 3000 User Manual

Capintec CAPTUS 3000
338 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #74 background imageLoading...
Page #74 background image
CAPINTEC, INC. CAPTUS
®
3000
Figure 5-15 Constancy Test Results Screen
An unacceptable deviation is indicated in red as HIGH.
If Deviation is indicated as HIGH, check the following:
• The positioning of the rod source, especially for the Probe detector.
• Ensure that there are no other sources (including dosed patients) in the area.
• Click the button or press Alt+T at the bottom of the Constancy Test screen.
Calibrated Activity and Calibration Date should match your Cs137 standard.
If a problem is found with any of the three items above, then correct the problem and
repeat the Constancy Test.
If no problems are found with the above items, then the efficiency of the Cs137 standard
should be re-measured. If this does not correct the problem, contact Capintec’s only
Authorized Service Center for service.
10. Click the button or press Alt+E to exit and return to Figure 5-1 Quality Assurance
Main Screen.
QUICK TIPS:
• To print current results for record keeping, click the button or press
Alt+P on Constancy Test screen after the test is completed.
• To view or print results for all the Quality Assurance tests performed to date,
click the button or press Alt+V (Refer to the
VIEWING AND
PRINTING HISTORY section on page 5-24).
5-14 QUALITY ASSURANCE July 14

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Capintec CAPTUS 3000 and is the answer not in the manual?

Capintec CAPTUS 3000 Specifications

General IconGeneral
BrandCapintec
ModelCAPTUS 3000
CategoryMedical Equipment
LanguageEnglish

Related product manuals