GE HEALTHCARE
DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
3 - 60 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x
3-9-14-2 DICOM Server IP Address Setting on the Scanner (cont’d)
7.) Select the IP-Address down-arrow to choose the Worklist Server from the pull-down menu.
NOTE: It is not possible to change the setting in the IP-Address field by editing it. See description
below, starting with step 8, to change the setting.
8.) Follow the steps below to change the IP-Address settings:
a.) From the Ip-Address pull-down menu, select <Modify> to display the IPs dialog.
b.) Select the server you want to modify from the IPs dialog and then select Modify.
Figure 3-55 DICOM Servers Properties dialog
Figure 3-56 Select <Modify>
Figure 3-57 Select the server to modify
DO NOT SELECT THE
SEARCH CRITERIAS BUTTON
IF NOT TOLD SO BY THE
ONLINE CENTER.
MODIFY
SELECT SERVER

Table of Contents