GE HEALTHCARE
DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
3 - 68 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x
3-9-14-6 DICOM Storage Properties Adjustments (cont’d)
Retry
Max # - The number of times the software will try to connect if the connection doesn’t succeed the
first time(s).
Interval - The interval in seconds between each attempt to connect.
Smiley
Select the Smiley button to ping the server. If the network connection is OK, the Smiley will smile, if the
connection fail, the Smiley will turn sad.
NOTE: The smiley checks if the remote server is accessible (ping). It is not a DICOM Echo (DICOM
ping), so it does not check AE title or port number.
Reopen per Image
If Reopen per image is enabled, Vivid 7 will create a new connection (association) for each image. This
may be useful for DICOM servers that do not accept different image types in the same association.
Storage Commitment
If you want to enable Storage Commitment, check the check-box and select the button “Storage
Commitment”.
In the dialog, enter IP address, Name, AE Title and Port Number.