GE HEALTHCARE
DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
Chapter 3 - System Setup 3 - 69
3-9-14-6 DICOM Storage Properties Adjustments (cont’d)
MPPS
If you want to enable MPPS, check the check-box and select the button “MPPS”. In the dialog, enter IP
address, Name, AE Title and Port Number.
Time-out
Adjust Time-out setting
Introduced with software versions v6.0, v5.2 and v4.2: The retry settings can be used to make jobs
retry on bad networks. There is no need to set retries for mobile (off-line) use when these software
versions are used.
Software v5.1.x/v4.1.x and lower: The retry settings for DICOM Storage are by default for mobile (off-
line) use. They ensure that jobs in the spooler will be sent automatically if the user reconnects within the
total retry period. If the customer does not have the mobile case, you can reduce the number of retries
and the interval. This will make the job fail faster if there is a network or server problem (and let the user
forward the job to another server if desired).
OK/Cancel
•Select OK to save all changes. The changes will take place after the scanner has been restarted.
•Select Cancel to return without saving any changes.
Figure 3-70 MPPS