3 - 84 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x
3-9-17-3      Query/Retrieve Verification (cont’d)
      Follow the Steps Below to Test the DICOM Query/Retrieve Workflow (cont’d)
- If the test fails, a pop-up dialog reporting the results as “0 of 1 OK” is displayed. In the example 
in Figure 3-86, the warning message pop up almost at once, and it states that for the Server 
Check (Ping), 1 of 1 is OK but the Device Check failed (0 of 1 OK). This indicates that the 
network connection functions OK, but the application on the remote server is not answering.
If the test fails directly, as described in this example, the AE Title is probably wrong.
A red “X” to the left of the Selected devices indicates that the test failed, but it doesn’t indicate the cause.
- If the Check button fails after a long time (corresponding to the timeout), the Warning will look 
like the example in Figure 3-88, where both tests failed.
In this case, the IP address or Port Number is probably wrong.
If Retrieve is Failing ...
If Retrieve is failing but Query is functioning, verify if the scanner has been registered as an receiver on 
the server.
Figure 3-86   Test (Check) Fails directly
Figure 3-87   Test (Check) Passed