GE HEALTHCARE
DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
3 - 96 Section 3-10 - Connectivity Setup - Software v3.x
3-10-8-2 DICOM Server IP Address Setting on the Scanner - Software Version v3.x (cont’d)
c.) Edit the name and/or the IP address of the server, see Figure 3-103.
d.) Select OK to save the new settings and close the dialog.
e.) Select OK to close the IPs dialog to return to the Properties dialog.
f.) Enter the DICOM server AE Title. This entry is case sensitive and must match exactly.
g.) Enter the DICOM server port.
h.) Select OK to close the Worklist properties dialog and save any changes.
Figure 3-103 Edit Name and/or IP address
Figure 3-104 DICOM Server’s Properties dialog
Name
IP address
AE Title
Port No