GE HEALTHCARE
DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
3 - 116 Section 3-11 - Connectivity Setup - Software v2.x
3-11-8 Verification of a Connection Set Up to EchoPAC PC - Software v2.x (cont’d)
• In the example in Figure 3-125, the red “X” next to the computer sign ( ), means that the check
failed. The cause may be;
- the setup of the network information, either on Vivid 7 or on the EchoPAC PC, or
- the network connection is broken somewhere between the Vivid 7 and the EchoPAC PC.
NOTE: If the Check button fails directly, the AE Title is probably wrong.
If the Check button fails after a long time (corresponding to the timeout), the IP address or Port
Number is probably wrong.
• A green light next to the computer sign means that a network connection is in place.
Figure 3-125 Verification of connection failed
Red “X”