DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
Chapter 3 - System Setup 3 - 121
3-11-8 Verification of a Connection Set Up to EchoPAC PC - Software v2.x (cont’d)
13.)In the list of “Services”, select “Remote Archive - Remote Storage”.
14.)In the list of Services select Remote Archive - Remote Storage.
15.)Change the Corresponding sql.ini name to reflect the name of the ECHOPAC7-nnnnnn
Workstation. The Corresponding sql.ini name is a combination of the host name plus a space and
the two letters “DB”.
Example: If the EchoPAC PC name is ECHOPAC7-987654, the Corresponding sql.ini name will be ECHOPAC7-
987654 DB.
16.)Change the Placement of Image Files and Placement of Report Files names so they reflect the
name of the EchoPAC PC Workstation.
Example: If the EchoPAC PC name is ECHOPAC7-987654, the Placement of Image Files and Placement of
Report Files are both ECHOPAC7-987654\Archive.
This completes the EchoPAC PC connection setup.
NOTE: Remember that these settings have no effect until you have logged on to the dataflow.
It may be necessary to log on and log off this dataflow before the settings work.
Figure 3-132 Dataflow services Choice and Setup
Remote Archive -
Remote Storage
Corresponding sql.ini
name
Storage location for
Image Files
Storage location for
Report Files

Table of Contents