DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
8 - 58 Section 8-9 - Software Loading Procedure - Software Version v2.x
8-9-7Move Images - Software Version v2.x (cont’d)
4.) Select Move Images from the More pull-down menu (Figure 8-51) to display the MOVE IMAGES
window (Figure 8-52).
5.) Adjust From date and To date to select the images from a defined time period.
6.) Check Keep days together if images from examinations performed on the same day are to be kept
on the same MO disk.
7.) Press Recalculate.
The Free disk space and Needed space information is updated. Make sure that the space needed
does not exceed the space available.
NOTE: If the space needed is bigger than the available space on the destination media, the time period
selected must be decreased by selecting an earlier date in the To date field.
Press Recalculate again and ensure that there is enough space.
8.) Select OK.
A progress indicator is displayed.
9.) After completion, Select OK.
10.)Remove MO-Disk with moved Images from MO-Drive.
11.)In case archive HD exceed MO Disk media capacity;
a.) Insert a new MO-Disk.
b.) Repeat step 5 through 11 as needed.
Figure 8-51 Select Move Images from the More pull-down menu
Figure 8-52 The Move Images Window
Move Images