8-14 PN: 10580-00343 Rev. H S820E UG
8-5 Table Display Format Vector Voltmeter
In the example shown in Figure 8-10 on page 8-15, measurement starts from
the default active Row1. When you press Enter to lock in the measurement
data for that row, the Green unlock icon changes to a Red lock icon to indicate
that the data in that row have been locked. The next available row (Row 2)
then becomes active with live data. Pressing Enter on Row 2 locks the data,
and the next available row (Row 3) then becomes active. This repeats until
the table has been completely filled. If you skip back to a previously filled row
and press Enter, then the pop-up message that is shown in Step 5 appears.
Pressing Continue clears the data for that row, and the row becomes active
with live data. After the live data are locked into that row, the next unused
row in the table (if any remain) will become live, and the standard sequence
returns.