Spaces in user names
If you allow spaces in user names, be aware of the following issues:
Leading and trailing spaces in user names are dicult to detect on the screen or
in printed reports.
The number of consecutive spaces in a user name is dicult to determine on the
screen or in printed reports.
Spaces in user names may cause confusion when searching for an audit or e-sig
record associated with a user name. To nd a record associated with a user name, you
must specify the user name exactly, including leading, consecutive, and trailing
spaces.
Set up messaging notifications
1.
In the Security screen (Figure 29), click Setup Messaging Notications to
display the Setup Notications dialog box.
2.
Select the ev
ents for notication:
Number (#) of failed authentications over specied time interval—A user
aempts to log in with an incorrect password. The message indicates the
number of failed authentications.
Session timeout for a user—No activity occurred in a user account for the
specied period of inactivity.
Account suspension for failed authentication—The user exceeds maximum
number of allowed failed authentications (login aempts with an incorrect
password).
Notication for SAE activation —Not supported.
Chapter 8 Use Security
, Audit, and E-Sig functions (SAE Module)
Administrators
8
194
3500/3500xL Genetic Analyzer User Guide—Data Collection Software v3.1

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 3500 and is the answer not in the manual?

Applied Biosystems 3500 Specifications

General IconGeneral
Number of capillaries8
Capillary Length50 cm
LaserSolid-state laser
Instrument TypeCapillary electrophoresis system
TechnologyCapillary Electrophoresis (CE)
Weight95 kg
Power Requirements100–240 V