Select objects to audit
1.
Select the objects and/or actions to audit.
Objects you can select for auditing (audit
records displayed in Object Audit
History):
Dye set
Size standard
Instrument protocol
PA protocol (primary analysis)
Assay
Plate template
File name convention
Results group
Plate
Sample files
Actions you can select for auditing (audit
records displayed in Action Log):
Export assay
Export plate record
Note: For a list of items that the system audits silently in addition to the
congurable items listed abov
e, see “Generate audit reports“ on page 204.
2.
Set the Audit Mode for each item you enable for auditing:
Prompt—The event is audited, a reason prompt is displayed, but the user
can cancel and continue without entering a reason.
Required—The event is audited, a reason prompt is displayed, and the user
must specify a reason.
Silent—The event is audited, no reason prompt is displayed.
3.
Click Save Seings.
Chapter 8 Use Security
, Audit, and E-Sig functions (SAE Module)
Administrators
8
202
3500/3500xL Genetic Analyzer User Guide—Data Collection Software v3.1