Select the actions that allow signature
IMPORTANT! Do not change e-sig seings during a spectral calibration.
By default, no events require electronic signature. To use e-sig, enable events that
require an e-signature.
1.
Select the checkbox next to an item in the E-Signature T
ype list to identify events
for which to allow electronic signature. This selection activates the E-Sig buon
for the selected items; it does require an electronic signature for these selections.
2.
(Optional) For each item that y
ou select:
a.
From the top-right of the screen, select a function after which the system will
prompt for electronic signature. This selection presents an e-sig prompt to
users when they perform a function. Users can sign or can continue without
signing.
b.
From the boom-right of the screen, select a function (for example, start run)
before which the system will check for required e-sigs. This selection
presents an e-sig prompt to users when they start a run if the required
signatures have not previously been made. Users must sign before they can
continue. For “check before” functions, you can also:
Change the number of signatures required.
Set a special authority for a signature: click the Authorities Required
eld, then select the user account or the user role to require for e-sig of
this function. By default, each required signature needs no special
authority; any user can sign.
Click Apply.
Chapter 8 Use Security, Audit, and E-Sig functions (SAE Module)
Administrators
8
212
3500/3500xL Genetic Analyzer User Guide—Data Collection Software v3.1

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 3500 and is the answer not in the manual?

Applied Biosystems 3500 Specifications

General IconGeneral
Number of capillaries8
Capillary Length50 cm
LaserSolid-state laser
Instrument TypeCapillary electrophoresis system
TechnologyCapillary Electrophoresis (CE)
Weight95 kg
Power Requirements100–240 V