A-10 Starling Monitor - User Guide
Miscellaneous Issues
Buzzer failure during
monitoring.
Cause:
Either
Buzzer turned Off.
Or
Buzzer malfunction.
If Buzzer turned Off
Access the Main Menu,
Session Controls &
Settings
options, then turn the Audible Alert
setting to On. See Figure 5-29 on page 5-34.
If Buzzer malfunction
Contact Baxter Service Center:
email: starling_rma@baxter.com
Tel (worldwide): +1 360-828-8685
Screen freeze Press and hold the Power On button on the Operator
Panel for a few seconds to reset the Monitor.
Device logo is displayed at
system start-up.
Blank screen or screen
display distorted, at
power On, indicates
malfunction.
Contact Baxter Service Center:
email: starling_rma@baxter.com
Tel (worldwide): +1 360-828-8685
Table A-4 Starling System Troubleshooting - Miscellaneous Issues (continued)
Issue Reason or Cause Solution