CAPINTEC, INC. CAPTUS
®
3000
To abort the procedure, click the
button or press the Y key. Collected data
will be discarded and the screen will return to Figure 10-45 Cr51 Blood Volume Test
Main Screen. To resume the test, click the
button or press the N key.
6. Next the screen appears as in Figure 10-61, prompting you to measure the plasma
standard.
Figure 10-61
To measure, place the plasma standard in the well and click the
button. To
quit counting, click the
button and Figure 10-60 will appear.
7. The screen will appear as shown in Figure 10-62, prompting you to measure the whole
blood sample.
Figure 10-62
To measure, place the whole blood sample in the well and click the
button.
To stop counting, click the
button and see Step 5 above.
8. The screen appears as shown in Figure 10-63, prompting you to measure the patient
plasma sample
Figure 10-63
November 13 LAB TESTS 10-41