EasyManua.ls Logo

Capintec CAPTUS 3000 - Page 252

Capintec CAPTUS 3000
338 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CAPINTEC, INC. CAPTUS
®
3000
Figure 11-18
Note: If calibration has not been performed, you will be prompted with the screen as
shown in Figure 11-19. After the calibration is complete, the program will return to
the Schilling Test measurement screen shown in Figure 11-18.
Figure 11-19
3. To start the measurement, click the
button or press Alt+S. The screen
appears as shown in Figure 11-20.
Figure 11-20
Beginning from here, the program leads you through a sequence of measurements.
Simply follow the on-screen instructions as explained below.
11-14 BIOASSAY November 13

Table of Contents

Related product manuals