EasyManuals Logo

Capintec CAPTUS 3000 User Manual

Capintec CAPTUS 3000
338 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #253 background imageLoading...
Page #253 background image
CAPINTEC, INC. CAPTUS
®
3000
a. Verify that there are no radioactive sources in the area and click the
button to measure background counts. The screen appears as shown in Figure 11-
21.
Figure 11-21
Note: Background counts must be measured everyday at least once, before
measuring any Staff Member. You may re-measure the background as often
as required. The last set of background data will be used to calculate net
activity.
b. The screen shown in Figure 11-22 appears next, prompting you to measure the Staff
Member.
Figure 11-22
To begin counting, click the button. The spectrum is displayed as shown
in Figure 11-21 above. To cancel measurements, click the
button. The
screen then appears as shown in Figure 11-23.
November 13 BIOASSAY 11-15

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Capintec CAPTUS 3000 and is the answer not in the manual?

Capintec CAPTUS 3000 Specifications

General IconGeneral
BrandCapintec
ModelCAPTUS 3000
CategoryMedical Equipment
LanguageEnglish

Related product manuals