CAPINTEC, INC. CAPTUS
®
3000
Figure 5-8
If you are performing the Constancy Test, Figure 5-9 will be displayed.
Figure 5-9
12. Place the Cs137 source rod in the selected detector. Since the Constancy Test is a
measure of reproducibility, it is important that the source be placed in exactly the same
position each time the Constancy Test is performed.
13. Click the button to proceed with the test, or else click the button
to stop the process. As the counting progresses, the live spectra and measurement times
are displayed.
14. To stop the counting, click the button or press Alt+A. The collected data is
discarded. To re-start the process, click the button or press Alt+S.
15. After the test is completed, the results of the test appear in the box below as shown in
Figure 5-10. Note the Deviation value at the bottom.
5-8 QUALITY ASSURANCE July 14