CAPINTEC, INC. CAPTUS
®
3000
3. Click the button or press Alt+M. The screen appears as shown in
Figure 6-6 with the name of the isotope at the top.
Figure 6-6
4. Click in the text box or use the Tab key to move around. Input the Calibration Activity,
Calibration Date and Calibration Time from the source label and Counting Time
desired in seconds.
5. To change the detector from PROBE to WELL, click in the appropriate radio button in the
Select Detector frame.
6. Click the
button or press Alt+N. The screen appears as shown in Figure 6-8.
Note: If calibration has not been performed for the current day for the selected detector,
you will be prompted with the screen as shown in Figure 6-7. If calibration is
performed now, the program will return to Figure 6-8 after calibration is complete.
Figure 6-7
6-6 ISOTOPES November 13