DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
Chapter 3 - System Setup 3 - 107
3-10-10-3 Query/Retrieve Verification (cont’d)
Follow the Steps Below to Test the DICOM Query/Retrieve Workflow
NOTE: This check uses both Ping and DICOM Ping.
- Ping is used to verify the TCP/IP connection to the server.
- DICOM Ping is used to verify that the DICOM application is answering
1.) Select QueryRetrieve from the Selected devices list
2.) Select Check.
- If the test pass, a pop-up dialog is displayed, see left illustration in Figure 3-114.
NOTE: If the Check button fails directly, the AE Title is probably wrong.
If the Check button fails after a long time (corresponding to the timeout), the IP address or Port
Number is probably wrong.
- Select OK to continue. A green check mark to the left of the selected devices indicates that
the test was successfully passed.
Figure 3-113 Check the DICOM Query/Retrieve workflow
Figure 3-114 Test (Check) Passed
QUERYRETRIEVE
CHECK