DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
3 - 108 Section 3-10 - Connectivity Setup - Software v3.x
3-10-10-3 Query/Retrieve Verification (cont’d)
Follow the Steps Below to Test the DICOM Query/Retrieve Workflow (cont’d)
- If the test fail, a pop-up warning is displayed, see left illustration in Figure 3-115.
Select OK to continue. A red “X” to the left of the Selected devices indicates that the test failed.
If Retrieve is Failing ...
If Retrieve is failing but Query is functioning, verify if the scanner has been registered as an receiver on
the server.
Figure 3-115 Test (Check) Failed