TRIGger Subsystem R&S FSL
1300.2519.12 6.282 E-11
C
haracteristics
*RST value: 625
SCPI: conform
M
ode
A–T
TRIGger<1|2>[:SEQuence]:VIDeo:LINE:NUMBer (option TV Trigger, B6)
With active TV trigger this command activates triggering at the horizontal sync signal of the
indicated line number.
Option TV Trigger, B6, is available from firmware version 1.10.
Parameter
<numeric_value>
Example
TRIG:VID:LINE:NUMB 17
Characteristics
*RST value: –
SCPI: conform
Mode
A–T
TRIGger<1|2>[:SEQuence]:VIDeo:SSIGnal:POLarity (option TV Trigger, B6)
With active TV trigger this command selects the polarity of the video sync signal.
Option TV Trigger, B6, is available from firmware version 1.10.
Parameter
NEGative | POSitive
Example
TRIG:VID:SSIG:POL NEG
Characteristics
*RST value: NEGative
SCPI: conform
Mode
A–T