R&S FSL  SENSe Subsystem (Power Meter, K9)  
1300.2519.12 6.383  E-11 
SENSe Subsystem (Power Meter, K9) 
The SENSe subsystem is organized in several subsystems. The commands of these subsystems 
directly control device–specific settings, they do not refer to the signal characteristics of the 
measurement signal. 
The SENSe subsystem controls the essential parameters of the analyzer. In accordance with the SCPI 
standard, the keyword "SENSe" is optional for this reason, which means that it is not necessary to 
include the SENSe node in command sequences. 
The following subsystems are included: 
• "SENSe:PMETer Subsystem" on page 6.383 
SENSe:PMETer Subsystem (Power Meter, K9) 
This subsystem controls the device settings for power sensor measurements. 
Commands of the SENSe:PMETer Subsystem 
– [SENSe<1|2>:]PMETer[:STATe] 
– [SENSe<1|2>:]PMETer:FREQuency 
– [SENSe<1|2>:]PMETer:FREQuency:LINK 
– [SENSe<1|2>:]PMETer:MTIMe 
– [SENSe<1|2>:]PMETer:MTIMe:AVERage[:STATe] 
– [SENSe<1|2>:]PMETer:MTIMe:AVERage:COUNt 
– [SENSe<1|2>:]PMETer:ROFFset[:STATe] 
– [SENSe<1|2>:]PMETer:TRIGger[:STATe] 
– [SENSe<1|2>:]PMETer:TRIGger:LEVel 
[SENSe<1|2>:]PMETer[:STATe] 
 
This command switches the power sensor measurements on or off. 
The numeric suffixes <1|2> are irrelevant. 
Parameter 
ON | OFF 
Example 
PMET ON 
Switches the power sensor measurements on. 
Characteristics 
*RST value: OFF 
SCPI: device–specific 
Mode 
PSM