Remote Commands of the 1xEV-DO BTS Analyzer Option (K84)  R&S FSL 
1300.2519.12 6.656  E-11 
SENSe Subsystem (EVDO, K84) 
The SENSe subsystem controls the essential parameters of the analyzer. In accordance with the SCPI 
standard, the keyword "SENSe" is optional for this reason, which means that it is not necessary to 
include the SENSe node in command sequences. 
The following subsystems is included: 
• "SENSe:CDPower Subsystem (EVDO, K84)" on page 6.656 
SENSe:CDPower Subsystem (EVDO, K84) 
This subsystem sets the parameters for the code domain measurements mode. The numeric suffix at 
SENSe<1|2> is irrelevant. 
Commands of the SENSe:CDPower Subsystem 
– [SENSe<1|2>:]CDPower:AVERage 
– [SENSe<1|2>:]CDPower:CODE 
– [SENSe<1|2>:]CDPower:ICTReshold 
– [SENSe<1|2>:]CDPower:IQLength 
– [SENSe<1|2>:]CDPower:CTYPe 
– [SENSe<1|2>:]CDPower:ICTReshold 
– [SENSe<1|2>:]CDPower:IQLength 
– [SENSe<1|2>:]CDPower:LEVel:ADJust 
– [SENSe<1|2>:]CDPower:MAPPing 
– [SENSe<1|2>:]CDPower:MMODe 
– [SENSe<1|2>:]CDPower:NORMalize 
– [SENSe<1|2>:]CDPower:OVERview 
– [SENSe<1|2>:]CDPower:PNOFfset 
– [SENSe<1|2>:]CDPower:QINVert 
– [SENSe<1|2>:]CDPower:SBANd 
– [SENSe<1|2>:]CDPower:SLOT 
– [SENSe<1|2>:]CDPower:TPMeas 
[SENSe<1|2>:]CDPower:AVERage 
 
This command can be precisely enabled by means of ON when the Code Domain Power 
analysis is active (refer to CALCulate<1|2>:FEED). If averaging is active, the CDP is calculated 
over all slots and displayed as called for by the 1xEV–DO Standard.  
Parameter 
ON | OFF 
Example 
CDP:AVER ON 
Activate averaging CDP relative over all slots and display on screen A.