Remote Commands of the 1xEV-DO BTS Analyzer Option (K84) R&S FSL
1300.2519.12 6.662 E-11
[
SENSe<1|2>:]CDPower:QINVert
This command inverts the sign of the Q–component of the signal.
T
he numeric suffix at SENSe<1|2> is irrelevant for this command.
Parameter
ON | OFF
Example
INST:SEL BC2K
Activates the 1XEV-DO BTS Analyzer option.
INIT:CONT OFF
Selects single sweep.
CDP:QINV ON
Activates the inversion of the Q–component sign.
INIT;*WAI
Starts measurement with synchronization.
Characteristics
RST value: OFF
SCPI: device-specific
Mode
EVDO
[SENSe<1|2>:]CDPower:SBANd
This command is used to swap the left and right sideband.
Parameter
NORMal | INVers
Example
CDP:SBAN INV
Swap sidebands.
Characteristics
RST value: NORMal
SCPI: device-specific
Mode
EVDO
[SENSe<1|2>:]CDPower:SLOT
This command selects the slot that is evaluated in the Code Domain Power Diagram and is
used for slot–based evaluations in the Code Domain Result Summary.
The numeric suffix at SENSe<1|2> is irrelevant for this command.
Parameter
0 to (capture length–1)
The capture length is defined via the [SENSe<1|2>:]CDPower:IQLength command.

Table of Contents