Tracking Generator (Models 13, 16 and 28)  R&S FSL  
1300.2519.12 4.134  E-11 
Tracking Generator (Models 13, 16 and 28) 
During operation the tracking generator emits a signal exactly at the input frequency of the R&S FSL.  
The tracking generator can be used in all measurement modes. Acquisition of test setup calibration 
values (see Source Cal  softkey) and normalization using these correction values (see Normalize 
softkey) is only possible in the tracking generator measurement mode. For details on measurement 
modes refer to "Measurement Mode Selection – MODE Key" on page 4.129. 
FFT filters (for details see "To choose the appropriate filter type" on page 4.19) are not available if the 
tracking generator is active. 
For measurements with running tracking generator it is recommended to set the start frequency to 3 x 
resolution bandwidth in order to meet the data sheet accuracy. 
Note:  The RF characteristics of some DUTs are especially sensitive concerning the input VSWR. In 
such cases insertion of 10–20 dB attenuation between the DUT and the tracking generator 
output is recommended. 
To open the tracking generator menu 
1. Press the MENU key. 
2. Press the Tracking Generator softkey. 
The tracking generator menu is displayed. 
Menu and softkey description 
– "Softkeys of the tracking generator menu" on page 4.137 
To display help to a softkey, press the HELP key and then the softkey for which you want to display 
help. To close the help window, press the ESC key. For further information refer to section "How to use 
the Help System". 
Further information 
– "Transmission measurement" on page 4.135 
– "Reflection measurement" on page 4.135 
– "Calibration mechanism" on page 4.136 
Tasks 
– To calibrate for transmission and reflection measurement