CALibration Subsystem R&S FSL
1300.2519.12 6.120 E-11
CALibration Subsystem
The commands of the CALibration subsystem determine the data for system error correction in the
instrument.
Commands of the CALibration Subsystem
CALibration[:ALL]?
CALibration:ABORt
CALibration:RESult?
CALibration:STATe
CALibration[:ALL]?
This command initiates the acquisition of system error correction data. A "0" is returned if the
acquisition was successful.
Note: During the acquisition of correction data the instrument does not accept any remote
control commands.
In order to recognize when the acquisition of correction data is completed, the MAV bit in the
status byte can be used. If the associated bit is set in the Service Request Enable Register, the
instrument generates a service request after the acquisition of correction data has been
completed.
Example
*CLS
Resets the status management.
*SRE 16
Enables MAV bit in the Service Request Enable Register.
*CAL?
Starts the correction data recording, and then a service request is generated.
Characteristics
*RST value: -
SCPI: conform
Mode
all