TRIGger Subsystem (WiMAX, K92/K93) R&S FSL
1300.2519.12 6.864 E-11
TRIGger Subsystem (WiMAX, K92/K93)
The trigger subsystem is used to synchronize device action(s) with events.
Commands of the TRIGger Subsystem
TRIGger[:SEQuence]:HOLDoff
TRIGger[:SEQuence]:MODE
TRIGger[:SEQuence]:LEVel[:EXTernal]
TRIGger[:SEQuence]:LEVel:POWer
TRIGger[:SEQuence]:LEVel:POWer:AUTO
TRIGger[:SEQuence]:HOLDoff
This command defines the length of the trigger delay. A negative delay time (pretrigger) can be
set in zero span only.
Parameter
–3.25 ms to 837.33 ms
Example
TRIG:HOLD 500us
A holdoff period of 500 µs is used after the trigger condition has been met.
Characteristics
*RST value: 0 s
SCPI: conform
Mode
OFDM, OFDMA/WiBro
TRIGger[:SEQuence]:LEVel[:EXTernal]
This command defines the level of the external trigger input for which triggering will occur.
Parameter
0.5 V to 3.5 V
Example
TRIG:MODE EXT
Sets the power trigger mode.
TRIG:LEV 1 V
Sets the triggering level to 1 V.
Characteristics
RST value: 1.4 V
SCPI: device–specific
Mode
OFDM, OFDMA/WiBro