R&S FSL TRIGger Subsystem (WiMAX, K92/K93)
1300.2519.12 6.865 E-11
TRIGger[:SEQuence]:MODE
This command configures how triggering is to be performed.
Parameter
IMMediate
No triggering is performed. This corresponds to the Free Run trigger mode.
EXTernal The next measurement is triggered by the signal at the external trigger input
e.g. a gated trigger.
POWer The next measurement is triggered by signals outside the measurement
channel.
Example
TRIG:MODE IMM
No triggering is performed.
Characteristics
*RST value: IMMediate
SCPI: device–specific
Mode
OFDM, OFDMA/WiBro
TRIGger[:SEQuence]:LEVel:POWer
This command sets the level of the input signal for which triggering will occur.
Parameter
–50 to 20 dBm
Example
TRIG:MODE POW
Sets the external trigger mode.
TRIG:LEV:POW 10 DBM
Sets the level to 10 dBm for RF measurement.
Characteristics
*RST value: –20 DBM