R&S FSL  DISPlay Subsystem (BLUETOOTH, K8)  
1300.2519.12 6.369  E-11 
DISPlay Subsystem (BLUETOOTH, K8) 
The DISPLay subsystem controls the selection and presentation of textual and graphic information as 
well as of measurement data on the display. 
Commands of the DISPlay Subsystem 
– DISPlay[:WINDow<1|2>]:TRACe<1...6>:Y[:SCALe]:PDIVision 
– DISPlay[:WINDow<1|2>]:TRACe<1...6>:Y[:SCALe]:RPOSition 
– DISPlay[:WINDow<1|2>]:TRACe<1...6>:Y[:SCALe]:RVALue 
DISPlay[:WINDow<1|2>]:TRACe<1...6>:Y[:SCALe]:PDIVision 
 
This command defines the scaling of the y–axis in the currently selected unit. The numeric suffix 
in TRACe<1...6> is irrelevant. 
Parameter 
<numeric_value> 
Example 
DISP:TRAC:Y:PDIV 10kHz 
Sets the Y scale to 10 kHz/div. 
Characteristics 
*RST value: – 
SCPI: conform 
Mode 
BT 
 
DISPlay[:WINDow<1|2>]:TRACe<1...6>:Y[:SCALe]:RPOSition 
 
This command defines the position of the reference value. In Bluetooth mode, the selected 
reference position is valid for all measurements. 
Characteristics 
*RST value: 50 PCT 
For further details refer to "DISPlay[:WINDow<1|2>]:TRACe<1...6>:Y[:SCALe]:RPOSition 
(models with tracking generator)" on page 6.136. 
 
DISPlay[:WINDow<1|2>]:TRACe<1...6>:Y[:SCALe]:RVALue 
 
This command defines the result value assigned to the reference position on the graticule.  
Example 
DISP:TRAC:Y:RVAL 0 
Sets the value assigned to the reference position to 0 Hz 
Characteristics 
*RST value: 0 Hz 
For further details refer to "DISPlay[:WINDow<1|2>]:TRACe<1...6>:Y[:SCALe]:RPOSition 
(models with tracking generator)" on page 6.136.