EasyManua.ls Logo

Rohde & Schwarz R&S FSL3 - Fetch Subsystem (Power Meter, K9)

Rohde & Schwarz R&S FSL3
1734 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
R&S FSL FETCh Subsystem (Power Meter, K9)
1300.2519.12 6.381 E-11
FETCh Subsystem (Power Meter, K9)
The FETCh subsystem contains commands for reading out results of complex measurement tasks.
The following subsystem is included:
"FETCh:PMETer Subsystem" on page 6.381
FETCh:PMETer Subsystem (Power Meter, K9)
The FETCh:PMETer subsystem contains commands to read measurement results of power sensor
measurements without starting the measurement itself.
Commands of the FETCh:PMETer Subsystem
FETCh<1|2>:PMETer?
FETCh<1|2>:PMETer?
This command reads the result of the power sensor measurement.
If no measurement has been performed, this command will lead to a query error. This command
is a query and therefore has no *RST value.
Example
FETC:PMET?
Reads the result of the power sensor measurement.
Characteristics
*RST value: –
SCPI: device–specific
Mode
PSM

Table of Contents

Related product manuals