R&S FSL  Measuring the Spectra of Complex Signals 
1300.2519.12 2.9  E-11 
Measurement example – Measuring the R&S FSL's intrinsic intermodulation 
Test setup: 
Signal
Generator 1
Signal
Generator 2
Coupler
[- 6 dB]
R&S FSL
 
Signal generator settings (e.g. R&S SMU): 
 
Level Frequency 
Signal generator 1  –4 dBm  999.7 MHz 
Signal generator 2   –4 dBm  1000.3 MHz 
Procedure: 
 
1.  Set the R&S FSL to its default settings.  
 Press the PRESET key. 
The R&S FSL is in its default state. 
 
2.  Set center frequency to 1 GHz and the frequency span to 3 MHz. 
 Press the FREQ key and enter 1 GHz.
 Press the SPAN key and enter 3 MHz.
3.  Set the reference level to –10 dBm and RF attenuation to 0 dB. 
 Press the AMPT key and enter –10 dBm.
 Press the RF Atten Manual softkey and enter 0 dB.
4.  Set the resolution bandwidth to 10 kHz. 
 Press the BW key. 
 Press the Res BW Manual softkey and enter 10 kHz.
The noise is reduced, the trace is smoothed further and the intermodulation products can be 
clearly seen. 
 Press the Video BW Manual softkey and enter 1 kHz.