R&S FSL Measuring the Spectra of Complex Signals
1300.2519.12 2.9 E-11
Measurement example – Measuring the R&S FSL's intrinsic intermodulation
Test setup:
Signal
Generator 1
Signal
Generator 2
Coupler
[- 6 dB]
R&S FSL
Signal generator settings (e.g. R&S SMU):
Level Frequency
Signal generator 1 –4 dBm 999.7 MHz
Signal generator 2 –4 dBm 1000.3 MHz
Procedure:
1. Set the R&S FSL to its default settings.
Press the PRESET key.
The R&S FSL is in its default state.
2. Set center frequency to 1 GHz and the frequency span to 3 MHz.
Press the FREQ key and enter 1 GHz.
Press the SPAN key and enter 3 MHz.
3. Set the reference level to –10 dBm and RF attenuation to 0 dB.
Press the AMPT key and enter –10 dBm.
Press the RF Atten Manual softkey and enter 0 dB.
4. Set the resolution bandwidth to 10 kHz.
Press the BW key.
Press the Res BW Manual softkey and enter 10 kHz.
The noise is reduced, the trace is smoothed further and the intermodulation products can be
clearly seen.
Press the Video BW Manual softkey and enter 1 kHz.

Table of Contents