EasyManua.ls Logo

Rohde & Schwarz R&S FSL3 - System Subsystem (Noise Figure, K30)

Rohde & Schwarz R&S FSL3
1734 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
R&S FSL SYSTem Subsystem (Noise Figure, K30)
1300.2519.12 6.563 E-11
SYSTem Subsystem (Noise Figure, K30)
This subsystem contains a series of commands for general functions.
Commands of the SYSTem subsystem
SYSTem:CONFigure:DUT:GAIN
SYSTem:CONFigure:DUT:STIMe
SYSTem:CONFigure:DUT:GAIN
This command specifies the maximum gain of the DUT.
Parameter
10 dB to 999.99 dB
Example
SYST:CONF:DUT:GAIN 10
Specifies the gain of the DUT to be 10 dB.
Characteristics
*RST value: 30 dB
SCPI: device–specific
Mode
NF
SYSTem:CONFigure:DUT:STIMe
This command spcifies the DUT settling time. It represents the time to wait for the DUT to settle
after a noise source has been turned on or off.
Parameter
0 to 20 s
Example
SYST:CONF:DUT:STIM 1000MS
Specifies a period of one second for the DUT to settle down after exposure to the noise source
has been removed.
Characteristics
*RST value: 50 milliseconds
SCPI: device–specific
Mode
NF