R&S FSL  SYSTem Subsystem (Noise Figure, K30)  
1300.2519.12 6.563  E-11 
SYSTem Subsystem (Noise Figure, K30) 
This subsystem contains a series of commands for general functions. 
Commands of the SYSTem subsystem 
– SYSTem:CONFigure:DUT:GAIN 
– SYSTem:CONFigure:DUT:STIMe 
SYSTem:CONFigure:DUT:GAIN 
 
This command specifies the maximum gain of the DUT. 
Parameter 
10 dB to 999.99 dB 
Example 
SYST:CONF:DUT:GAIN 10 
Specifies the gain of the DUT to be 10 dB. 
Characteristics 
*RST value: 30 dB 
SCPI: device–specific 
Mode 
NF 
 
SYSTem:CONFigure:DUT:STIMe 
 
This command spcifies the DUT settling time. It represents the time to wait for the DUT to settle 
after a noise source has been turned on or off. 
Parameter 
0 to 20 s 
Example 
SYST:CONF:DUT:STIM 1000MS 
Specifies a period of one second for the DUT to settle down after exposure to the noise source 
has been removed. 
Characteristics 
*RST value: 50 milliseconds 
SCPI: device–specific 
Mode 
NF