R&S FSL Status Reporting System
1300.2519.12 5.27 E-11
Reset Values of the Status Reporting System
Table 513 contains the different commands and events causing the status reporting system to be
reset. None of the commands, except *RST and SYSTem:PRESet, influences the functional instrument
settings. In particular, DCL does not change the instrument settings.
Table 5–13 Resetting the status reporting system
Event Switching on
supply voltage
DCL,SDC
Power–On–
Status–Clear
(Device
Clear,
Selected
Device
Clear)
*RST or
SYSTem:PRE
Set
STATus:PRE
Set
*CLS
Effect 0 1
Clear STB,ESR
yes
yes
Clear SRE,ESE
yes
Clear PPE
yes
Clear EVENt parts of
the registers
yes
yes
Clear ENABle parts
of all OPERation and
QUEStionable
registers; Fill ENABle
parts of all other
registers with "1".
yes
yes
Fill PTRansition parts
with "1";
Clear NTRansition
parts
yes
yes
Clear error queue yes yes
yes
Clear output buffer yes yes yes 1) 1) 1)
Clear command
processing and input
buffer
yes yes yes
1) Every command being the first in a command line, i.e., immediately following a <PROGRAM
MESSAGE TERMINATOR> clears the output buffer.