DISPlay Subsystem (Analog Demodulation, K7) 
The DISPLay subsystem controls the selection and presentation of textual and graphic information as 
well as of measurement data on the display. 
Commands of the DISPlay Subsystem 
– DISPlay[:WINDow<1|2>]:SIZE 
– DISPlay[:WINDow<1|2>]:TRACe<1...6>:Y[:SCALe]:PDIVision 
– DISPlay[:WINDow<1|2>]:TRACe<1...6>:Y[:SCALe]:RPOSition 
– DISPlay[:WINDow<1|2>]:TRACe<1...6>:Y[:SCALe]:RVALue 
– DISPlay[:WINDow<1|2>]:TRACe<1...6>:Y:SPACing 
DISPlay[:WINDow<1|2>]:SIZE 
 
This command switches the measurement window for active analog demodulation to full screen 
or half screen. 
For further details refer to "DISPlay[:WINDow<1|2>]:SIZE" on page 6.130. 
 
DISPlay[:WINDow<1|2>]:TRACe<1...6>:Y[:SCALe]:PDIVision 
 
This command defines the scaling of the y–axis in the current unit. The numeric suffix in 
TRACe<1...6> is irrelevant. The numeric suffixes <1|2> are irrelevant. 
 
Parameter 
<numeric_value> 
Example 
DISP:TRAC:Y:PDIV +10kHz 
Sets the Y scale to 10 kHz/div. 
Characteristics 
*RST value:  
20 PCT  linear AM display 
50 kHz  linear FM display 
2 rad linear PM display 
10 dB  logarithmic AF spectrum display 
 
SCPI: conform 
Mode 
ADEMOD