Measurements on Modulated Signals  R&S FSL  
1300.2519.12 2.30  E-11 
The following figure shows the passband characteristics of the 30 kHz channel filter in the R&S FSL. 
Fig. 2-23  Frequency response of the 30 kHz channel filter for measuring the power in the 2000 MC1 
adjacent channel 
Measurement example 2 – Measuring adjacent channel power of a W–CDMA uplink 
signal 
Test setup: 
Signal
generator
R&S FSL
 
Signal generator settings (e.g. R&S SMU): 
Frequency: 1950 MHz 
Level:  4 dBm 
Modulation:  3 GPP W–CDMA Reverse Link  
Procedure: 
 
1.  Set the R&S FSL to its default state. 
 Press the PRESET key. 
The R&S FSL is in its default state. 
 
2.  Set the center frequency to 1950 MHz. 
 Press the FREQ key and enter 1950 MHz.