R&S FSL  SENSe Subsystem  
1300.2519.12 6.173  E-11 
[
SENSe<1|2>:]AVERage[:STATe<1...6>] 
 
This command switches on or off the average calculation for the selected trace (STATe<1...4>). 
T
he numeric suffixes <1|2> are not relevant.  
Parameter 
ON | OFF 
Example 
AVER OFF 
Switches off the average calculation for trace 1. 
AVER:STAT3 ON 
Switches on the average calculation for trace 3. 
Characteristics 
*RST value: OFF 
SCPI: conform 
Mode 
all 
 
[SENSe<1|2>:]AVERage:COUNt 
 
This command defines the number of measurements which contribute to the average value. 
It should be noted that continuous averaging will be performed after the indicated number has 
been reached in continuous sweep mode. 
In single sweep mode, the sweep is stopped as soon as the indicated number of measurements 
(sweeps) is reached. Synchronization to the end of the indicated number of measurements is 
only possible in single sweep mode.  
The [SENSe<1|2>:]AVERage:COUNt command effects the same as the 
[SENSe<1|2>:]SWEep:COUNt command. In both cases, the number of measurements is 
defined whether the average calculation is active or not.  
The number of measurements applies to all traces.  
The numeric suffixes <1|2> are not relevant.  
Parameter 
0 to 32767 
Example 
SWE:CONT OFF 
Switching to single sweep mode. 
AVER:COUN 16 
Sets the number of measurements to 16. 
AVER:STAT ON 
Switches on the calculation of average. 
INIT;*WAI 
Starts the measurement and waits for the end of the 16 sweeps. 
Characteristics 
*RST value: 0 
SCPI: conform 
Mode 
all