R&S FSL SENSe Subsystem
1300.2519.12 6.173 E-11
[
SENSe<1|2>:]AVERage[:STATe<1...6>]
This command switches on or off the average calculation for the selected trace (STATe<1...4>).
T
he numeric suffixes <1|2> are not relevant.
Parameter
ON | OFF
Example
AVER OFF
Switches off the average calculation for trace 1.
AVER:STAT3 ON
Switches on the average calculation for trace 3.
Characteristics
*RST value: OFF
SCPI: conform
Mode
all
[SENSe<1|2>:]AVERage:COUNt
This command defines the number of measurements which contribute to the average value.
It should be noted that continuous averaging will be performed after the indicated number has
been reached in continuous sweep mode.
In single sweep mode, the sweep is stopped as soon as the indicated number of measurements
(sweeps) is reached. Synchronization to the end of the indicated number of measurements is
only possible in single sweep mode.
The [SENSe<1|2>:]AVERage:COUNt command effects the same as the
[SENSe<1|2>:]SWEep:COUNt command. In both cases, the number of measurements is
defined whether the average calculation is active or not.
The number of measurements applies to all traces.
The numeric suffixes <1|2> are not relevant.
Parameter
0 to 32767
Example
SWE:CONT OFF
Switching to single sweep mode.
AVER:COUN 16
Sets the number of measurements to 16.
AVER:STAT ON
Switches on the calculation of average.
INIT;*WAI
Starts the measurement and waits for the end of the 16 sweeps.
Characteristics
*RST value: 0
SCPI: conform
Mode
all